Temperature significantly affects conductivity (EC) and TDS measurements, as detailed below:
1. Effect of Temperature on Conductivity (EC) Measurements
Conductivity (EC) is a measure of a solution's electrical conductivity, and its magnitude is closely related to factors such as the concentration and migration rate of ions in the solution. Temperature's effect on conductivity is primarily manifested in the following ways:
Increased ion thermal motion: As temperature rises, ion thermal motion intensifies, accelerating ion migration. This makes it easier for ions to conduct current under the influence of an electric field, resulting in an increase in solution conductivity.
Decreased solvation: As temperature increases, the solvation degree of ions decreases, the ion solvation radius decreases, and conductivity increases accordingly.
Decreased solution viscosity: Increasing temperature decreases solution viscosity, facilitating ion movement and thus increasing conductivity.
Change in dielectric constant: Increasing temperature decreases the dielectric constant of the bulk solution, increasing interionic forces and decreasing conductivity. However, experiments have shown that the rate of increase in electrolyte solution conductivity with increasing temperature is comparable to the rate of decrease in viscosity with increasing temperature.
II. The Effect of Temperature on TDS Measurement
TDS (total dissolved solids) refers to the amount of dissolved solids in a solution and is typically measured using an approximate conversion relationship between conductivity (EC) and TDS. Because temperature significantly affects conductivity, it can also indirectly affect TDS measurements.
Direct Measurement Deviation: TDS meters based on conductivity infer TDS by measuring the conductivity of a solution. If the temperature changes during the measurement process, the solution's conductivity will change accordingly. The TDS meter may mistake this change in conductivity for a change in TDS, resulting in measurement deviation. For example, in real-world applications, if the seawater temperature rises from 20°C to 30°C and the TDS meter is not temperature compensated, the measured TDS value may be higher than the actual TDS value.
Calibration Curve Deviation: TDS meters are typically calibrated before shipment based on the conductivity-TDS relationship at a specific temperature, generating a calibration curve. However, when the actual measurement temperature differs from the calibration temperature, the conductivity-TDS relationship will deviate from the calibration curve, resulting in inaccurate measurement results.
III. Temperature Compensation Measures
To minimize the effects of temperature on conductivity (EC) and TDS measurement results, a temperature compensation function is typically configured in the instrument. Common temperature compensation methods include:
Manual Adjustment: Calibrate the conductivity meter by immersing it in constant-temperature water (e.g., 25°C). Manual adjustments are then made based on the measurement results to ensure the meter correctly identifies and compensates for temperature effects.
Automatic Adjustment: Modern conductivity meters are generally equipped with automatic temperature compensation. Simply insert a temperature sensor into the sample to perform temperature compensation. This can typically be configured within the instrument's settings interface.
Online Temperature Compensation: For complex samples, real-time temperature compensation is required. This can typically be achieved by connecting the conductivity meter and a temperature sensor to perform online temperature measurement and compensation, resulting in more accurate test results.
Through appropriate temperature compensation adjustments, the effects of temperature on conductivity (EC) and TDS measurements can be eliminated, resulting in more accurate and reliable test data.